AutoREID

Description

The ASTM D323, D4953, ISO 3007, IP69 and other similar test methods cover procedure for the determination of vapor pressure of gasoline, volatile crude oil, and other volatile petroleum products, components and feedstocks.
This method is suitable for testing air-saturated liquid samples that exert an air-saturated vapour pressure between 9,0 kPa and 180,0 kPa at 37,8 °C.
This test method called Reid method is part of many petroleum product specifications. The method is particularly suitable for the determination of vapor pressure of crude oil for transportation, storage and general handling purpose.
The AutoREID apparatus from AD systems is a modern automated instrument in full compliance with ASTM D323-B and similar standards.
The instrument directly measures the Reid vapor pressure (RVP) without correlation or calculation.
Compact instrument with two measurement positions allows performing two simultaneous measurements independently.
An additional position in the bath is dedicated for air chamber conditioning to increase the test productivity.
The AutoREID design allows testing relatively viscous and sticky samples or samples containing mechanical sediments without a risk of the instrument damage.
All parts of the test pressure vessel are easy to disassemble and clean.
Vigorous mechanical agitation accelerates pressure stabilization even on "difficult" samples like crude oil.

Operation:
- After the test vessel is connected by a quick connector the measurement starts automatically.
- The pressure vessel is agitated by reverse rotation and the pressure curve is displayed in real time on the large graphic screen.
- As soon as the maximum stable pressure is reached the test stops automatically and saturated vapor pressure (RVP) is recorded.

Reports:
- All details related to each test result are memorized.
- Complete test report including sample ID, operator name, date/time, RVP value, pressure curve is displayed on the screen and saved into non-volatile memory.
- It can be printed, transferred on a USB memory stick and/or send to a LIMS when the AutoREID is connected to a LAN.
- Multi-criteria search function facilitates retrieval of test report in the built-in database.

Quality and Safety:
- To assure high quality of results and safety of operation the watchdog system permanently monitors the bath temperature stability, for pressure leaks, bath liquid level, overpressure or overheating.
- Transparent bath cover helps for visual inspection of the pressure vessels during the test.
- To assure high quality of results and safety of operation the watchdog system permanently monitors the bath temperature stability, for pressure leaks, bath liquid level, overpressure or overheating.
- Transparent bath cover helps for visual inspection of the pressure vessels during the test.
- Instrument calibration is easy to perform on-site via a dedicated menu.
- Pressure transducers can be checked against a reference manometer at any time.
- The bath temperature can be checked with reference thermometer. The bath cover has a special opening for this purpose.
- All information related to the instrument calibration is memorized in the journal to ensure a perfect traceability

Brand

AD Systems

AD Systems specializes in the design, manufacturing, selling and servicing of unique test equipment for the evaluation of physical properties and performances of fuels and lubricants. All of the products offered by AD Systems were developed in response to specific needs within the petroleum industry. AD Systems has in the past and is currently working with governmental agencies and research departments of petroleum companies to develop prototypes and subsequently to manufacture various instruments and systems. AD Systems is an active member of the American Society for Test and Materials (ASTM®) as well as with Energy Institute, GFC and BNPé.